Dynaflow MK5 Eco Manuel - Page 7

Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} Dynaflow MK5 Eco. Dynaflow MK5 Eco 8 pages.

FUNCTION
1.
"Fire Alarm"
2.
"Spare"
3.
"Remote Alarm"
4.
"Remote Start"
5.
"Low Level"
6.
"Isolator 1"
7.
"Pressure Alarm"
8.
"Isolator 2"
9.
"Low Air Flow"
10. "Battery Alarm"
11. "Mains Alarm"
12. "Pump Configuration"
13. "Syncroflow"
14. "Spray Bar"
15. "GPO Outlets"
16. "Lighting"
17. "Aux Relay"
18. "Alarm Relay"
19. "Sash Input"
20. "Air Sensor"
21. Commissioning"
22. "Auto Start"
23. "RS485"
24. "Remote Panel"
25. "PH Meter"
26. "Air Con"
27. "Sash Warning"
28. "Dual Sash"
29. "Green Sash"
30. "BMS Interface"
31. "Green Lighting"
32. "Service Meter"
33. "Pump 1"
34. "Pump 2"
35. "Valve 1"
36. "Valve 2"
37. "VSD Direct"
38. "FAN Disable"
39. "AUTO SASH"
40. "Services"
Dynasafe
ECO Management Control System
Configuration Functions
ALARM
AL1
External alarm
AL2
Not Used
AL3
Used for external alarm
AL4
Used for external start
AL5
Scrubber low level
Emergency Isolator
Measure duct press
AL6
External isolator
AL7
Low air flow indicator
Low battery alarm
Mains Failure
Scrubber pump (240v output)
Variable Volume Sash Control
Fume cupboard spray bar
Fume cupboard GPO's
Fume cupboard light
Control board relays
General alarm relays
Require to calibrate sash(s)
Variable Hot wire airflow sensor
Activate commissioning
Auto start after power failure
BMS interface (High Level)
Additional control panel
Monitor PH Control
Start/Stop output
Exceed sash opening
Allows twin sash to operate
Pre-set alarm to close sash
BMS Interface
1.
Power Failure
2.
Start
3.
Low Level
4.
Airflow Failure
5.
Emergency Stop
Turn light "OFF" at set Time
Set Next Service Schedule
Waste System only
Waste System only
Waste System only
Waste System only
Adjust VSD or Control Valve from the Fume Cupboard
Deactivates "FAN" Button, FC Runs Continually
Automatic Sash Movement
Auto start after power failure
40 Programmable Functions
INPUTS/OUTPUTS
(volt free input) NO Relay
(volt free input) NO Relay
(volt free input) NO Relay
(volt free input) NO Relay
(volt free input) NO Relay
(Front Panel)
N/A) 4 to 20
(volt free input) NO Relay
(volt free input) NO Relay
(240v output)
(Output to Contactor) 240 Volt
(N/A)
(setup)
(Allows Services)
N/A
(D/S Fume Cupboards)
(alarms only)
(NO or NC 240v)
(Duel Sash)
(Low Level) outputs Volt Free
(Fire Mode)
(NO Services)