iHealth Neo BP5S Manuel de l'utilisateur - Page 6
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} iHealth Neo BP5S. iHealth Neo BP5S 14 pages.
SPECIFICATIONS
1. Product name: iHealth Neo Wireless Blood Pressure Monitor
2. Model: BP5S
3. Classification: Internally powered; Type BF applied part; IP22, No AP or APG; Continuous operation
4. Machine size: approx. 5.39"x 2.22"x 0.65" (137mm×56.5mm×16.5mm)
5. Cuff circumference: 8.66" to 16.54" (22cm-42cm) and 16.54" to 18.90"(42cm-48cm) (Optional)
6. Weight: approx. 8.47 oz (240g) (including cuff)
7.Measuring method: Oscillometric method, automatic inflation and measurement
8.Memory volume: 200 times with time and date stamp (off-line measurement only)
9. Power: DC:5.0V
1.0A, Battery: 1*3.7V
Li-ion 800mAh
10. Measurement range:
Cuff pressure: 0-300 mmHg
Systolic: 60-260 mmHg
Diastolic: 40-199 mmHg
Pulse rate: 40-180 beats/minute
11. Accuracy:
Pressure: ±3 mmHg
Pulse rate: ±5%
12.Wireless communication:
Bluetooth V4.1 Class 2
Frequency Band: 2.400-2.4835 GHz
13. Environmental temperature for operation: 5°C-40°C (41°F -104°F)
14. Environmental humidity for operation: ≤85%RH
15. Environmental temperature for storage and transport:-20°C-55°C (-4°F-131°F)
16. Environmental humidity for storage and transport: ≤90%RH
17. Environmental pressure: 80kPa-105kPa
18. Battery life: more than 130 measurements on a full charge
19. The blood pressure measurement system includes accessories: pump, valve, cuff, LED screen and sensor.
Note: These specifications are subject to change without notice.
GENERAL SAFETY AND PRECAUTIONS
1. Read all of the information in the User Guide and other provided instructions before operating the unit.
2. Consult your physician for any of the following situations:
a) The application of the cuff over a wound or inflamed area.
b) The application of the cuff on any limb with intravascular access or therapy, or an arteriovenous(A-V) shunt.
c) The application of the cuff on the arm on the side of a mastectomy.
d) Simultaneous use with other medical monitoring equipment on the same limb.
e) The blood circulation of the user needs to be checked.
3. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
4. Blood pressure measurements determined by this product are equivalent to those obtained by professional
healthcare practitioners using the cuff/stethoscope auscultation method within the limits prescribed by the American
National Standard, Electronic or Automated Sphygmomanometer.
6 / 14