GaleMed 2151 Manuel d'instruction - Page 7

Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} GaleMed 2151. GaleMed 2151 11 pages.

Performance Characteristics 
The performance characteristics for MR‐100™Plus    Manual Resuscitators 
will vary from user to user depending on a variety of factors: ambient 
temperature, patient lung compliance, ventilator frequency &      size of 
operator's hands. The following data has been derived from MR‐100™ Plus   
Manual Resuscitators in accordance with ISO standards. 
 
Frequency range 
 
o
o
o
-18
C/ 0
22
C/72
Cycle rate 
Cycle rate 
Adult 
38 
45 
Child A 
92 
105 
Child B 
72 
78 
Infant A 
95 
95 
Infant B 
60 
60 
These results were obtained under the following conditions: 
Adult: VT- 600ml, Compliance -0 02L/cmH
Child A: VT- 70ml, Compliance -0 01L/cmH
Child B: VT- 300ml, Compliance -0.02 L/cmH
Infant A: VT- 20ml, Compliance -0.001 L/cmH
Infant B: VT-70ml, Compliance -0 01 L/cmH
 
 
 
o
o
o
50
C/ 122
Cycle rate 
45 
105 
76 
95 
60 
O, Resistance - 20cmH
O/L/s   
2
2
O, Resistance- 20cmH
O/L/s   
2
2
O, Resistance- 20cmH
O/L/s 
2
2
O, Resistance- 400cmH
O/L/s   
2
2
O, Resistance- 20cmH
O/L/s 
2
2
Specifications 
Storage temp:
- 4 0 ' C / - 1 0 ' F to 6 0 ' C / 140'F
 
Operating temp:
- 1 8 ' C / O'F to 5 0 ' C / 1 2 2 ' F
 
 
Materials:
Mask, Duckbill valve, Mushroom
valve, Relief valve seal, Bag body,
Oxygen reservoir bag, Oxygen
Silicon rubber:
reservoir bag band & plug.
 
Non-rebreathing valve housing,
Polysulfone:
Pressure relief valve housing, A l l - in
one intake valve housing, Pressure
 
relief valve stem & cap.
Poly-vinyl-ch loride:
Oxygen tubing
 
Stainless steel:
Pressure relief valve
 
spring.
Aluminum:
Pressure relief n a m e plate.
 
E P D M :
O-ring
 
Connections:
 
Patient port
15 m ml D/ 22 m m O D
Silicon bag neck
24 m ml D
 
Oxygen gas inlet
6 m m O D
Intake valve port
2 5 m m O D
 
Reservoir bag port
2 5 m m l D