Performance Characteristics
The performance characteristics for MR‐100™Plus Manual Resuscitators
will vary from user to user depending on a variety of factors: ambient
temperature, patient lung compliance, ventilator frequency & size of
operator's hands. The following data has been derived from MR‐100™ Plus
Manual Resuscitators in accordance with ISO standards.
Frequency range
o
o
o
-18
C/ 0
F
22
C/72
Cycle rate
Cycle rate
Adult
38
45
Child A
92
105
Child B
72
78
Infant A
95
95
Infant B
60
60
These results were obtained under the following conditions:
Adult: VT- 600ml, Compliance -0 02L/cmH
Child A: VT- 70ml, Compliance -0 01L/cmH
Child B: VT- 300ml, Compliance -0.02 L/cmH
Infant A: VT- 20ml, Compliance -0.001 L/cmH
Infant B: VT-70ml, Compliance -0 01 L/cmH
o
o
o
F
50
C/ 122
F
Cycle rate
45
105
76
95
60
O, Resistance - 20cmH
O/L/s
2
2
O, Resistance- 20cmH
O/L/s
2
2
O, Resistance- 20cmH
O/L/s
2
2
O, Resistance- 400cmH
O/L/s
2
2
O, Resistance- 20cmH
O/L/s
2
2
Specifications
Storage temp:
- 4 0 ' C / - 1 0 ' F to 6 0 ' C / 140'F
Operating temp:
- 1 8 ' C / O'F to 5 0 ' C / 1 2 2 ' F
Materials:
Mask, Duckbill valve, Mushroom
valve, Relief valve seal, Bag body,
Oxygen reservoir bag, Oxygen
Silicon rubber:
reservoir bag band & plug.
Non-rebreathing valve housing,
Polysulfone:
Pressure relief valve housing, A l l - in
one intake valve housing, Pressure
relief valve stem & cap.
Poly-vinyl-ch loride:
Oxygen tubing
Stainless steel:
Pressure relief valve
spring.
Aluminum:
Pressure relief n a m e plate.
E P D M :
O-ring
Connections:
Patient port
15 m ml D/ 22 m m O D
Silicon bag neck
24 m ml D
Oxygen gas inlet
6 m m O D
Intake valve port
2 5 m m O D
Reservoir bag port
2 5 m m l D