table 2 for transient electromagnetic immunity
guidAnCe And mAnufACturer's deCLArAtiOn –
eLeCtrOmAgnetiC immunitY
The GRPro
2.1 is intended for use in the electromagnetic environment specified below. The
customer or the user of the GRPro
2.1 should assure that it is used in such an environment.
immunitY tEst
iEC 60601 tEst lEvEl
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
Voltage dips, short
<5% U
T
interruptions and
(>95% dip in U
)
T
voltage variations
for 0.5 cycle
on power supply
40% U
T
input lines
(60% dip in U
)
T
IEC 61000-4-11
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the AC mains voltage prior to application of the test level.
T
EN- 16
ComplianCE lEvEl
ElECtromagnEtiC EnvironmEnt – guidanCE
± 6 kV contact
Floors should be wood, concrete or
± 8 kV air
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
± 2 kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
environment.
± 1 kV for input/
output lines
± 1 kV line(s) to
Mains power quality should be that
line(s)
of a typical commercial or hospital
environment.
± 2 kV line(s) to
earth
<5% U
Mains power quality should be that
T
(>95% dip in U
)
of a typical commercial or hospital
T
for 0.5 cycle
environment. If the user of the GRPro
2.1 requires continued operation
40% U
T
during power mains interruptions, it is
(60% dip in U
)
T
recommended that the GRPro
for 5 cycles
powered from an uninterruptible power
70% U
T
supply or a battery.
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
table 3 for rf electromagnetic immunity
guidAnCe And mAnufACturer's deCLArAtiOn – eLeCtrOmAgnetiC
The GRPro
customer or the user of the GRPro
immunitY tEst
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
2.1 be
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the GRPro
above, the GRPro
observed, additional measures may be necessary, such as re-orienting or relocating the GRPro
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
immunitY
2.1 is intended for use in the electromagnetic environment specified below. The
2.1 should assure that it is used in such an environment.
iEC 60601 tEst lEvEl
ComplianCE lEvEl
3 Vrms
3Vrms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
2.1 is used exceeds the applicable RF compliance level
2.1 should be observed to verify normal operation. If abnormal performance is
ElECtromagnEtiC EnvironmEnt – guidanCE
Portable and mobile RF communications
equipment should be used no closer to any
part of the GRPro
2.1, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
recommended separation distance
d = 1.2 √P 80 MHz to 800 MHz
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
2.1.