      4.  Lightly tap the table with one finger. If the readings
     vary more than 0.5 mg after tapping, the table is 
     probably too unstable for accurate weight readings 
     during a test and will need to be made more rigid 
     or another location found for this equipment. 
                 Contact your AquaLab distributor if the scale 
     readings are unstable after placing the unit on a  
     stable structure.
Desiccant Cartridge
Desiccant Receiver
Black Valve
Black Receiving Hole
   •  Desiccant tubes are installed by orienting the tube so that
     the black colored valve inserts into the black receiving hole
     as shown in Figure 2. Before insertion make sure that the 
      desiccant tube lid is firmly sealed onto the desiccant tube. 
      (See Manual for instructions) Lift the tube into the area 
      above the desiccant receiver and then insert the tube into 
      the mating holes while shaking the tube slightly. Once in 
      place, the desiccant tube is sealed for flow purposes and 
      may move side to side if contacted.
Figure 2: Desiccant installation
V S A Q u i c k S t a r t G u i d e