Centurion Scientific C2006 Manuel d'instruction - Page 6
Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Centurion Scientific C2006. Centurion Scientific C2006 6 pages. Led benchtop centrifuges
OPEn
- You may now open the Lid.
NOTE: If
druc H
t
t
Test
ES
(all LED models)
This allows you to turn on or off the buzzer and to check
the sensors for, Tachometer, Rotor recognition, Lid open
& Imbalance detector.
See 23 Each go 0/1 as you turn rotor or operate Imbalance detector or shut lid.
Centrifuge lid in open position.
Press function (keep held down) whilst turning on at rear switch. When test shows release function button.
Press time up arrow See numbers in time display go to .............. 18
Buzzer off or on (0/1)................................................................................... 18
Set speed Rpm digits in 10 or 100 Rpm steps.................................. 19
EEtE St -
E prom check Press Speed up arrow..................................................... Number not shown
8888 88 -
LED display check increment speed up arrow.................................... Number not shown
Rotor rec Tacho Lid open Imbalance........................................ 24
Turn rotor Turn rotor Close lid Push rotor to one side..................... Display changes from 0 to 1
Bu2t St -
Buzzer test press speed up arrow........................................................... Number not shown
Do9t St -
Watchdog test Press speed up arrow.
diagonally
opposite and of the same weight.
Check if the tubes are broken If so use better quality, or if glass
or
drIuE
(drive) show please Contact Service Department, or refer to technical manual.
rotor? -
settings, it will have readjusted. Please recheck your speed and try
again.
If this continues contact the service department.
rrEc ? -
contact the service department.
Routine
Mains off at rear switch.
UUAIt -
Wait 5 seconds press Lid open
tAcho? -
long and fouling the lid or anything else is
stopping rotation of the rotor.
-
An imbalance has occurred Please check the tubes are
The sensor has found the speed to be incorrect to the rotors
A problem with the sensor or inverter has occurred please
HELL 0 -
(Hello) should display, then back to normal.
Check that your tubes are not too