Fault codes. Ambient model
Fault codes are shown in the Speed section of the LED display
The following List show what each Code means and what to do.
F A 1 0
No rotation seen
F A 2 1
Imbalance BG
F A 2 2
Imbalance BD
F A 2 3
Imbalance MG
F A 2 4
Imbalance MD
F A 2 5
Imbalance base sensor
F A 2 6
Imbalance Motor sensor
F A 2 7
Data time out
F A 28
Imbalance
F A 3 0
Lid is open
F A 4 0
Rotor recognition
F A 50
Drive
F A 51
Druc
*
Obround viewing port
shown.
at the rear of the Centrifuge
*
Situated
Check the Following with the Lid open.
Tubes are too tall and foul the Lid, replace with correct ones
Check the rotor to see if it freely turns, if not look for
Objects in the bowl that may be fouling and remove.
If not Contact our Service Department
Check the Following.
Weigh your tubes to 0.1g of each other
Place tubes in Rotor opposite each other
(see previous Rotor pages for instructions)
If Fault code persists call our Service Department.
As above.
As above.
As above.
Contact our Service Department.
Contact our Service Department.
Contact our Service Department.
See the following Rotor pages
Close the Lid, press down on 2 front corners, push firmly.
problem persists Contact our Service Department
Wrong rotor fitted or missing magnets if so.
Contact our Service Department
Contact our Service Department. Also See below
Contact our Service Department. Also see below
Look at the Obround viewing port (Rear)
For FA 50 and FA 51
Note the Code shown in the inverter display.
Turn off the power at the wall mains switch only.
WAIT 3 minutes, Turn on again and Run. If problem persists
Contact our Service Department, with the Inverter code