IMPORTANT
Check the Centrifuge and the Packaging for any shipping damage.
Inform the shipping company and Centurion Scientific immediately
Check the order for completeness, if not contact Centurion Scientific
Location
The Centrifuge should only be operated indoors.
The set up location must fulfil the following requirements
A safety zone of at least 30Cm must be maintained around the Centrifuge
People and hazardous substances must be kept out of this zone
during centrifugation
The bench or supporting structure must be stable and free from resonance
The Bench or supporting structure must be suitable for horizontal use of the
centrifuge
The Centrifuge should Not be exposed to heat or strong sunlight
Ultra violet rays can damage the stability of plastics and coatings
Do not subject the Polypropylene, Delrin rotors or accessories to direct
sunlight
The Zone location must be well ventilated at all times
Store the original packaging, Keep this for future service to base , if required.
Transporting the Centrifuge
• Always remove the Rotor before transportation
• Due to its weight the Centrifuge should be carried by several people
• Transport the Centrifuge upright and if possible in its packaging
• Always lift the centrifuge at both sides (see below). NOT front to rear.
Read before using
2