IMPORTANT
Check the Centrifuge and the Packaging for any shipping damage.
•
Inform the shipping company and Centurion Scientific immediately
Check the order for completeness, if not contact Centurion Scientific
•
Location
The Centrifuge should only be operated indoors.
The set up location must fulfil the following requirements
A safety zone of at least 30Cm must be maintained around the Centrifuge
•
People and hazardous substances must be kept out of this zone
•
during centrifugation
The bench or supporting structure must be stable and free from resonance
•
The Bench or supporting structure must be suitable for horizontal use of the
•
centrifuge
The Centrifuge should Not be exposed to heat or strong sunlight
•
Ultra violet rays can damage the stability of plastics and coatings
•
Do not subject the Polypropylene, Delrin rotors or accessories to direct
•
sunlight
The Zone location must be well ventilated at all times
•
Store the original packaging, Keep this for future service to base , if required.
•
Transporting the Centrifuge
• Always remove the Rotor before transportation
• Due to its weight the Centrifuge should be carried by several people
• Transport the Centrifuge upright and if possible in its packaging
• Always lift the centrifuge at both sides (see below). NOT front to rear.
Read before using
2