Gamry Interface 5000 Instructions - Page 2
Parcourez en ligne ou téléchargez le pdf Instructions pour {nom_de_la_catégorie} Gamry Interface 5000. Gamry Interface 5000 2 pages. Installing standard cable on reference 600+
Également pour Gamry Interface 5000 : Manuel de démarrage rapide (4 pages)
Place the calibration cell inside the calibration shield as depicted in Figure 2. Please make sure to then
place the lid on the calibration shield and connect the black lead of the cell cable to the grounding post
on the side of the calibration shield.
Figure 2: Proper orientation of calibration cell inside calibration shield.
Running Calibration
Once the calibration cell is properly connected inside the grounded calibration shield, you may run
calibration as normal. Start the Gamry Framework software and select the Experiment menu, then
select the Utilities sub-menu, and finally click on Calibration. Once the calibration dialog box is shown,
you will simply select the instrument and select Both as the calibration type as shown in Figure 3.
Figure 3: Sample of calibration dialog box.
Press OK to begin calibration. Once the calibration completes, your instrument is now ready to run.
2