When the CAPBs is switched on, the measuring instrument
performs zero calibration. Zero calibration must be performed in
COfree air. If you want to take a measurement, verify that you only
switch on the measuring instrument in CO-free air.
If zero calibration is performed in air with a CO concentration of >
50 ppm, the sensor does not yield measured values, but
generates an error message. New zero calibration in CO-free air
clears this error.
INCORRECT MEASURED VALUES
Perform zero calibration immediately after switching on in
COfree fresh air.
Failure to follow these instructions can result in death, serious injury
or equipment damage.
Up to 45 hours of battery operation