11. Activate the recovery pump as with a normal operation. The combination of the nitrogen pressure above the
filter and the vacuum pressure below the filter will cause the solution to flow through the media. Depending on
media type and solution viscosity flow may vary. It is recommended to suspend the addition of heat to the
collection vessel until ample solution has passed through the filter allowing a liquid solution to pool in the
collection vessel. This will help prevent burning of extract materials if evaporation/recovery is happening at a
rate faster than the flow through the filter.
12. Once the solution is finished passing through the filter (as observed by sight glasses when available), you may
vent off the nitrogen from the system to keep it from circulating through the system and being recovered into
the recovery tank.
Disposal
13. Once the run is complete, remove the filter cartridge from the housing assembly and discard.
*Some contents of filters may cause irritation if handled. Wear appropriate protective gloves and eye wear when
handling. Do not remove contents of filters as some contents may be harmful to respiratory system. In case of
accidental spillage, use appropriate respiratory masks. Do not eat contents. Not intended for internal use.