Alicat Scientific EPCD Series Manuel d'utilisation - Page 12

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Alicat Scientific EPCD Series. Alicat Scientific EPCD Series 17 pages. Pressure vacuum controller

Alicat Scientific EPCD Series Manuel d'utilisation
Sending Setpoints as Integers
In this method, your controller's full scale range (FS) is represented by a value
of 64000, and a zero setpoint is represented by 0. To calculate your intended
setpoint, use the following formula:
Integer value = 64000 ×
Example 1:
A desired setpoint of +5.44 inH2OG on a 10-inH2OG pressure
controller is calculated as 64000 × 5.44 ⁄ 10.00 = 32704. The command to
assign the setpoint based on this integer value is:
New setpoint:
Example:
Example 2:
When using a bidirectional pressure controller, 0 represents -100%
of full scale, 32000 represents 0, and 64000 represents +100% of full scale.
Use the formula below to calculate the integer value.
Integer value = 64000 ×
A desired setpoint of +5.44 inH
calculated as 64000 × [5.44 + 10]/10.00 = 9882. The command to assign the
setpoint based on this integer value is:
New setpoint:
Example:
Example 3:
A setpoint of -5.44 inH
calculated as:
Integer value = 64000 x (-5.44 + 10)/10 = 29184
[desired setpoint]
[unit ID][setpoint as integer where 64000 is FS]
a32704
(setpoint of 5.44 inH2OG)
[desired setpoint + FS]
OG on a 10-inH
2
[unit ID][setpoint as integer where 64000 is FS]
a9882
(setpoint of 5.44 inH2OG)
OG on the same bidirectional controller is
2
12
[device FS]
[device FS]
OG pressure controller is
2