TRADITIONAL PUSH BUTTON CALIBRATION
To calibrate at 10%: Push and Hold the Cal Button (About 1 Second) until confir-
mation blinks, then release the button. The output will read 10% of full scale after
calibration.
To calibrate at 25%: Push and Hold Cal Button (About 5 Seconds) until you see
two sets of confirmation blinks. Then release the button. The output will read
25% of full scale after calibration. (If no confirmation blink occurs, inadequate
calibration weight may have been used)
After calibration: Remove the weight and observe the output. It should read
0 VDC or 4 mA with nothing touching the tension sensing roller.
Once calibrated, tension data is also available over the network connec-
tion, however it should be noted that the calibration performed can be considered
unitless and rangeless unless the value in the CalRange Register and the CalUnits
register at calibration was valid. In this case the TENSION_P may be the most
desirable tension register. See Accessing Tension Data for more information.
WEB PAGE INTERFACE CALIBRATION
See Manual for instructions.
CALIBRATION USING THE ETHERCAT® INTERFACE
ESI file can be found on the product web page,
see Manual for instructions.