Specifications (General):

FEATURE

Selector Switch Modes Available

Regulated Vacuum Range

Maximum Flow Regulator only
(nominal)

Safety Relief Valves

Source Vacuum Range

Vacuum Gauge

Range
Size
Type
Accuracy ±3% of full scale

Materials of Construction

Case
Knobs
Internal Parts
Metal Parts

Operating Temperature

Operating Humidity

Intermittent OFF Time
Intermittent ON Time
Int. OFF Time Range
Int. ON Time Range

Overall Dimensions

Weight

"N/A" means not available on this model Vacutron suction regulator.
*
**
Maximum flow is affected by vacuum source and fittings. Values shown are without fittings.
MODE
Continuous
Intermittent
Off
0-110 mm Hg  10
Intermittent
0-130 mm Hg  10
Continuous
Cont. (Special Model)
10 Year Warranty
Intermittent
Continuous
Intermittent
Continuous
Continuous
Cont. (Special Model)
12 to 30 in Hg (305 to 762 mm Hg)
(40.6 to 101.3 kPa)
Unit calibrated at 18 in Hg Vacuum Source
0 to 150 mm Hg (Increments of 5 mm Hg)
0 to 20 kPa (Increments of 2 kPa)
2 Inch Diameter Gauge Face
Diaphragm
High Impact Polycarbonate
Soft Thermoplastic Rubber Outer Shell Molded on a Thermoplastic Body
Polycarbonate, Acetal Resin, Silicon, and Fluorocarbon Rubber
Brass, Stainless Steel and Plated Steel
50F to 90F (10C to 32C)
35% to 96% Relative Humidity (RH) Non-condensing
Factory set to 7-12 seconds at 18 in Hg Vacuum Source
Factory set to 15-20 seconds at 18 in Hg Vacuum Source
5-30 seconds (field adjustable ) at 18 in Hg Vacuum Source
5-30 seconds (field adjustable ) at 18 in Hg Vacuum Source
6.8 Inches (174 mm) High x 2.7 Inches (67 mm) Wide x 6.0 Inches (152 mm) Deep
1.7 lbs. (0.79 kg)
PEDIATRIC
INTERMITTENT
CONT/INT
Yes
Yes
Yes
0-150 mm Hg  10
0-14.7 kPa  1.3
0-20 kPa  1.3
0-17.3 kPa  1.3
2.5 LPM
7-9 LPM **
40 LPM
110 mm Hg  10
150 mm Hg  10
0-14.7 kPa  1.3
0-20 kPa  1.3
130 mm Hg  10
0-17.3 kPa  1.3
0 to 300 mm Hg (Increments of 5 mm Hg)
0 to 40 kPa (Increments of 2 kPa)
2 Inch Diameter Gauge Face
Diaphragm
5
CONT/INTER
No
Yes
Yes
Yes
Yes
Yes
0-150 mm Hg  10
0-20 kPa  1.3
0-320 mm Hg  20
N/A *
0-42.5 kPa  2.7
0-320 mm  20
0-42.5 kPa  2.7
7-9 LPM **
N/A *
65 LPM
150 mm Hg  20
20 kPa  2.7
310 mm Hg  20
N/A *
41.3 kPa  2.7
310 mm Hg  20
41.3 kPa  2.7