DH Instruments MPG1 WITH PG7302 Fiche d'instruction
Parcourez en ligne ou téléchargez le pdf Fiche d'instruction pour {nom_de_la_catégorie} DH Instruments MPG1 WITH PG7302. DH Instruments MPG1 WITH PG7302 2 pages. Interconnection kit for mpg1 with rpm4 or pg7302
INTERCONNECTION KIT FOR MPG1 WITH
RPM4 OR PG7302, P/N 402265, INCLUDES:
DESCRIPTION
Nipple, 6 in.
(152 mm), DH500
Nipple, 12 in.
(305 mm), DH500
Nipple, 5 in. x 1/8 in.
(127 mm x 3 mm),
DH500 tips
Tee DH500
Plug DH500/DH700
Adaptor, DH500F x
1/8 in. (3 mm) NPTM
INTRODUCTION
This kit provides the tubing and fittings necessary to
connect an MPG1 pressure controller to either an
RPM4 reference pressure monitor or to a PG7302
piston gauge, and also to a device under test. This
document contains instructions for connecting the
fittings and tubing.
INSTALLATION OF INTERCONNECTIONS
Remove Plugs from MPG1 and RPM4 or
PG7302: Unscrew the glands from the test ports
of the MPG1 and RPM4 or PG7302. Remove the
collar and plug from inside each gland.
should be two plastic plugs and one stainless steel
plug. The plastic plugs will not be used. Keep the
stainless steel plug.
Install glands and collars onto flexible 5 in.
nipples: Slide a gland over a nipple so that the
threads of the gland face outward. Next, screw a
collar onto the nipple by turning it to the LEFT.
Allow two nipple threads to show between the
collar and nipple tip.
RPM4 – Install glands and collars onto the 6 in.
a
nipple.
PG7302 – Install glands and collars onto the
b
12 in. nipple.
PRESSURE
PART
QTY
RATING
500 MPa
1
103975
(72 500 psi)
500 MPa
1
103976
(72 500 psi)
200 MPa
2
123019
(30 000 psi)
400 MPa
1
100291
(60 000 psi)
700 MPa
1
100285
(100 000 psi)
200 MPa
2
102819
(30 000 psi)
MPG1 with RPM4 or PG7302
For
Install nipples into three way tee: Install the 5 in.
nipple into the center port of the three way tee.
Install either the 6 in. or 12 in. nipple into either of the
end ports on the three way tee as shown in Figure 1.
NO.
Tighten the nipples with a 5/8 in. wrench or crescent
wrench. The thick nipple on the right connects to the
MPG1. The thin nipple in the center of the plug
connects to the RPM4 or to the PG7601. In Figure 1
the left port is currently plugged, but the remaining
thin 5 in. nipple can go in the left port and connect to
the DUT.
Figure 1: Two nipples installed in 3-way tee.
Optional: Extend the front feet on RPM4: The front
feet of the RPM4 can be extended so that the RPM4
is inclined for easier reading and display access.
There
RPM4 A70M or less - install NPT adaptor into
RPM4.
(10 000 psi) or less have 1/8 in. NPTF test ports.
These ports require the use of one of the DH500F
by 1/8 in. NPTM adaptors. Wrap Teflon
tape around the NPT threads of the adaptor three
times, and install the adaptor into the RPM4.
Install the free end of the center nipple into
RPM4 or PG7302:
should be inserted into the test port of the RPM4 or
PG7302. Screw in the gland loosely by hand. The
nipple can be bent slightly to ensure that nipple
enters straight into the test port. Do not tighten the
gland in the RPM4 or PG7302 yet, since it will be
tightened when the system is primed.
Interconnection Kit
P/N 402265
Instruction Sheet
RPM4s that have ranges of 70 MPa
The center, thin, 5 in. nipple
®
thread