Maintenance

Cleaning the spray arms

The spray arm jets can get clogged.
Check daily and clean if necessary.
^ Use a sharp pointed object to push
particles into the spray arm jets and
rinse thoroughly under running water.
^ Remove the mobile unit.
34
Labwasher spray arms
^ Loosen the nut with a 60 mm wrench
and pull the spray arm upwards or
downwards respectively.
Mobile unit / insert spray arms
^ Turn the nut with the spray arm
counterclockwise and pull the spray
arm downwards.
^ Refit the spray arms after cleaning in
the reverse order.
After replacing the spray arms
check that they rotate freely.