The ASPEED2 aspirator is a professional device suitable for home and outpatient use, specific for secretion-aspiration. It is
equipped with Vacuum regulator (1), vacuum gauge (4), 1000 cc pots. with protection device, for the liquid intake
on the suction pump, that interrupts the suction flow. It is lubrication-free, handy, easy to use, reliable, sturdy and quiet. The
ASPEED 2 aspirator is supplied with the following 3A accessories: 1000 cc. pot with protection device, power supply cable,
connection tube in sterilisable silicon Ø 6x12 short, connection tube in sterilisable silicon Ø 6x12 long, sterile and single
usepipe, sterile and single use manual regulator, single use bag, Single use antibacterial filter.
Note: Use original 3A accessories ONLY.
This is a medical device and it must be used only upon medical prescription. It must be used as described in this instruction ma-
nual. The user must read and understand in full the information regarding the use and maintenance of the unit. For any question,
do not hesitate to contact your preferred retailer. MICROBIAL CONTAMINATION: due to pathologies with risk of microbial
infection and contamination a personal use of the accessories is recommended (ask your physician). **
The manufacturer does anything possible to provide a product equipped with the highest quality and safety features, although,
being it an electric device, some fundamental safety directions must be observed:
• Kids and non self-sufficient people must always use the device under strict supervision of a competent adult in his full-capaci
ty and who has read and understood this manual.
• This device must only be used for its intended purpose: aspirator for outpatient and domestic use; any other use is conside
red improper and dangerous and the manufacturer cannot be held responsible for any consequence of misuse.
• Never use adapters for voltages different from the voltage indicated on the data label on the back of the device.
• Keep the cable away from hot surfaces.
• This device is not suitable for use in presence of anaesthetic blend inflammable with air, oxygen, or nitrous oxide.
• Never handle the power supply cable plug with wet hands, and never use the device in the bath tub or in the shower. **
• Never leave the device near water, never immerse it in any liquid, never wet it, if it gets wet immediately unplug it from
the power supply socket before touching it. Do not use the device if the plug or the power supply cable are worn or wet
(immediately send it to your preferred retailer).
• The device is not watertight.
• Always unplug the power supply cable immediately after use.
• Maintenance and/or repair must be carried out only by authorised personnel. Non-authorised repairs void the warranty.
• Do not turn over the pot while it is connected to the device in function, since the liquid would come into contact with the
hydro phobic antibacterial filter, immediately blocking suction; in case this happens, empty the pot and replace the
antibacterial filter.
• The device is equipped with a safety fuse, mounted inside the device, in case you wish to replace it, unplug the device
before proceeding.
• The pipe and the manual control of the flow are sterile and single-use products: they must be replaced after each applica
tion. Check the expiration date on the original packaging of the pipe and the suction flow manual control and
verify the integrity of the sterile packaging.
• The antibacterial filter is a single-use device and it must be replaced after each application.
• Before each use always perform the cleaning and disinfecting operations as mentioned in the paragraph "Cleaning and
disinfecting operations" of this instructions manual.
Before each use, verify that all the accessories are perfectly clean
in accordance with the instructions of the section "CLEANING AND DISINFECTING OPERATIONS".
. Connect the device as shown in figure.
With the vacuum regulator (1) you can set the depression value (bar). depression - check Rotate the knob towards "+" to
increase the vacuum and rotate it towards "-" to decrease it, the values can be read on the vacuum gauge (4).
Turn the device on by putting the switch on the "I" position (ON) (5).
After the application turn off the device and unplug the cable form the power supply socket and carry out the cleaning and
disinfecting operations as described in the "CLEANING AND DISINFECTING OPERATIONS" paragraph.
The 1000 cc collection pot provided with the aspirator can be used in two ways. as sterilisable collection pot, or as
collection pot with single use bag.