PEEP (Positive end expiratory pressure)
PEEP can be fitted to the DPA 02 by connecting to the expiratory valve as shown in the picture
below ensuring the correct direction of flow.
Fig. 7
To adjust the PEEP valve
The PEEP valve can be removed by pulling the valve from the clear case. To adjust the valve,
turn red cap clockwise to increase pressure and anti-clockwise to reduce pressure. The PEEP
valve pressure ranges from 0-20 cmH20.
Fig. 8