Aquatec ocean e-vip Manuel d'utilisation - Page 14

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Aquatec ocean e-vip. Aquatec ocean e-vip 16 pages.

Problem
Possible
causes
Button for emer-
Rechargeable bat-
gency lowering
tery is flat
does not function
Cable not insert-
ed all the way
Connection plug
board not at-
tached
Button is faulty
Seat frame is
Motors were
tipped
raised to different
heights
Carrying out a reset:
1.
Press the following three buttons on the hand con-
trol simultaneously for ca. 3 seconds:
Red STOP button (5, fig. 12)
Forward button (3, fig. 12)
UP button (8, fig. 12)
A beep is heard while the three buttons are pressed.
2.
When the indicator lamp "SERVICE" (6, fig. 12)
goes out, release the buttons. The device is ready
for use again.
3.
Carry out a functional inspection. To do this, move
the chair in all directions.
4.
Then lower the drive mechanisms for height ad-
justment to the lower end position until they audi-
bly disconnects.
5.
Move the drive mechanism for the tilt adjustment
into the rear end position until it audibly discon-
nects.
Important!
If the error occurs again after reset, replace the
device or the faulty party and send them in.
EN
Remedy
Recharge the bat-
tery (see
section 5.2) or use
a replacement
Insert cable all the
way
Attach the con-
nection plug board
(clicks in)
Carry out a reset
(see below)
Press the DOWN
button on the hand
control (2, fig. 12)
until the drive
mechanisms are
completely low-
ered and you hear
the click sound of
the end switch
7

Care and maintenance

Important!
Hygiene is especially important as regards
the toilet. Regularly clean the commode
and keep it clean.
This product can be disinfected. Choose
the disinfectant in accordance with AQUA-
TEC care guidelines. These are available
on request from AQUATEC.
Regularly check that the screws are firmly
fitted.
The commode is maintenance-free provid-
ed you follow the operating and care in-
structions. Regular, recurrent safety inspec-
tions are required.
7.1

Cleaning the seat frame and main frame

CAUTION!
There is a risk of injury from electric shock.
Before cleaning, close all caps and protect
all plugs.
Important!
Do not use abrasive cleaning agents.
Do not detach the holders for the accesso-
ries.
1.
Close all caps and protect all plugs.
2.
Take off any fitted accessories.
3.
Pull the footrests out of the guides.
4.
Take off the seat plate and the backrest cover.
5.
Wash the individual parts with commercial deter-
gents (e.g. all purpose cleaning agents) using a
cloth or brush.
6.
Rinse the parts under warm water.
7.
Dry the parts with a cloth.
8.
Reassemble the commode.
Important!
Clean the backrest cover at a maximum tem-
perature of 60 °C in a commercial washing ma-
chine.
CAUTION!
Burns
Let the commode cool down before touch-
ing it.
In clinics, in addition to the cleaning operations de-
scribed above, the commode should be cleaned at
regular intervals at 85 °C for 3 minutes in the auto-
clave.
23