Acculab VI-1200 Manuel d'instruction - Page 17

Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Acculab VI-1200. Acculab VI-1200 20 pages.

Acculab VI-1200 Manuel d'instruction
Category
Ambient Conditions
Stability Range
External CAL Units
External Calibration
Auto Zero
Baud Rate
Parity
Stop Bits
Handshake
Print Stream
Printed Output
Menu Codes Reset
Note: Settings 5--, 6-- and 7-- have no effect on VI-Series balances because they affect
performance of the RS-232 port included with VIR-Series balances only.
page 16
User Configurable Internal Menu Codes
Segment Position
L
C
R
Value
1
1
1
very stable
1
1
2*
stable (default)
1
1
3
unstable
1
1
4
very unstable
1
3
1
⁄ digit
1
3
2
fi digit
1
3
3
1 digit
1
3
4*
2 digits
1
3
5
4 digits
1
4
1*
grams (default)
1
5
1*
external calibration is enabled (default)
1
5
2
external calibration is disabled
1
6
1*
on (default)
1
6
2
off
5
1
1
150
5
1
2
300
5
1
3
600
5
1
4*
1200 (default)
5
1
5
2400
5
1
6
4800
5
1
7
9600
5
2
1
mark
5
2
2
space
5
2
3*
odd (default)
5
2
4
even
5
3
1*
1 stop bit (default)
5
3
2
2 stop bits
5
4
1
software
5
4
2
hardware, w/2 chars after CTS (default)
5
4
3*
hardware, w/1 char after CTS
6
1
1
manual print, regardless of stability
6
1
2*
manual print when stable (default)
6
1
3
continuous print, regardless of stability
6
1
4
continuous print, when stable
7
1
1*
LCD contents only (default)
7
1
2
header and LCD contents
9
1
1
reset balance to factory defaults
*=Factory default setting.
TECHNICAL
APPENDICES