7 CALIBRATION
All SMART weighing scales come pre-calibrated from the factory. However the
scale may require re-calibration. Temperature changes, geographic gravity
variations, altitude changes and abuse are few reasons why a scale may need
recalibration. The scales can be calibrated using a known external weight.
EXTERNAL USER CALIBRATION
7.1
STEP
To start calibration turn the scale off and
then turn it on. During the count down,
1
press and hold down the Zero button (do
not release until the display shows "CAL"
Make sure there is no weight on the pan.
Press Zero / Enter key and the display
2
shows "noLoAd" momentarily and shows
the counts at zero level.
Pause for 2 seconds at the previous step
and press Zero/Enter key. The display
3
shows "LoAd" momentarily and shows the
internal counts.
Put a known weight on the weighing pan
(Ex.: 200g). Press the Enter key. The
4
display will now show all "000" with the
units digit blinking.
If you put 200g weight, Enter "200".
Change the blinking digit by pressing the
5
mode button. Shift the changed digit to
left by pressing the "P" button once.
Press the Enter key once again. The display
6
will show "......" for some time and then
show the weight
PROCEDURE
13
NANO Series Precision Balance
DISPLAY