Weighing in Percent
Purpose
This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight.
Example: Determine an unknown percentage; store the weight on the balance/scale as the reference percentage (100%)
Step
1. Select application program
2. Select Weighing in Percent
3. Confirm setting
Symbol " " on the display: application is active
4. Place empty container on the balance/scale
5. Tare the balance
6. Place the reference weight for 100% on the
balance/scale
7. If desired, change the number of decimal places
displayed: 100.0%, 100.00% or 100% (factory setting) (F) repeatedly
8. Confirm selected no. of decimal places
9. Place unknown weight on the balance/scale
10. Toggle display between weight and percentage
11. Unload the balance/scale
12. Weighing in Percent application:
clear the reference percentage
13. Reactivate Weighing in Percent (if no other
application program has been selected)
14. Repeat procedure starting from Step 5.
8
www.balances.com is the original deep discount distributor
Key (or instruction)
(ZERO) > 2 sec
(F) repeatedly
(ENTER)
(ZERO)
(ENTER)
(F) repeatedly
(ENTER) > 2 sec
(F)