TROUBLESHOOTING GUIDE

Under certain conditions, the performance of the camera may be adversely affected by factors other than defects.
If problems occur, please review the following list and take remedial action as needed. If you cannot solve the
problem after checking the entire list, please contact Olympus for assistance.
Problem
a) The PC will not start up.
b) Software will not start up.
c) Live image is not displayed.
d) Still images cannot be recorded.
e) Picture is too bright.
25
Cause
The PCI interface board is not inserted
completely into the PC's PCI slot.
The camera and PC are connected
improperly.
The PCI interface board is inserted into
the PC's PCI slot improperly.
The PCI interface board is not connected
to the ATX power supply.
The device driver is installed improperly.
The camera and PC are connected
improperly.
The microscope illumination is off.
The microscope is not set to the camera
light path.
The illumination or specimen focusing
is adjusted improperly.
The sensitivity or exposure time is set
improperly.
The camera and PC are connected im-
properly.
The DP Controller is processing record-
ing.
The DP Manager is processing file save,
etc.
The PC memory is insufficient.
Exposure correction is set in the +
direction.
The metering area is set to a dark area
outside the region of interest.
AE lock, which was set when the expo-
sure time was longer than the currently
required exposure time, is active.
Remedy
Turn off the PC, unplug the power cord
and insert the PCI interface board into
the PCI slot of the PC securely.
Turn off the PC and connect the
camera and PCI interface board
properly using the interface cable.
Turn off the PC, unplug its power cord
and insert the PCI interface board into
the PC's PCI slot.
Turn off the PC, unplug its power cord
and connect the PCI interface board
to the ATX power supply properly.
Install the device driver.
Turn off the PC and connect the
camera and PCI interface board
properly using the interface cable.
Turn on the microscope illumination,
adjust the lighting and focusing
correctly, and select the camera
light path.
Set the sensitivity, exposure mode, ex-
posure time and level properly.
Turn off the PC and connect the
camera and PCI interface board
properly using the interface cable.
Wait until the recording processing
completes before starting recording
of the next image. In certain cases, it
may be required to press the Cancel
button of the progress bar and record
the image again.
Wait until the processing completes
before starting recording of the next
image.
Close the image being displayed
on the DP Manager or close other
software before retrying recording.
Return the exposure correction value
to 0 and set the desired exposure
correction value.
Move the metering area to the area
where you want to obtain optimum
exposure.
Cancel AE lock.
Page
7
7/9
7
7
10-14
7/9
(Online
manual)
7/9
(Online
manual)
(Online
manual)
(Online
manual)
(Online
manual)
(Online
manual)