AccuLine 40-6535 Manuel d'instruction - Page 13

Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} AccuLine 40-6535. AccuLine 40-6535 20 pages. Automatic-leveling horizontal rotary laser level

AccuLine 40-6535 Manuel d'instruction
X-direction self-check with the same method as Y-direction
self-check, and mark point C and point D by turns.
7. If point C and point D are within 1/32" at 50', the accuracy is
within tolerance. Otherwise reference section 12 of this
document.
Enter self-calibration
In power off status, press the power and the tilt buttons of the laser unit
simultaneously. Then let go of the power button while still pressing
the tilt button. 10 seconds later, the self-calibration indicator lights
of X direction and Y direction flash 3 times simultaneously. Let go of
the tilt button, the instrument enters self-calibration mode. Remote
control is necessary for self-calibration.
X Direction Self-calibration
Select self-calibration direction
Make X direction of the instrument face to the reference line. Press
X/Y button on the remote control once to select X direction for
self-calibration. The X direction indicator light in the "CALIBRATOR"
area of the instrument keypad is lighted as shown in the following
figure, and the rotator head begins to rotate.
©2007 Johnson Level & Tool
13