Fig. 6
7
5
Mounting the Intermediate Attachments
@
1. Using the Allen screwdriver, loosen the observation tube clamping screw
@ on the microscope frame.
2. Insert the circular dovetail mount at the bottom of the intermediate
attachments into the opening on the microscope frame and clamp
by tightening the clamping screw @.
When Using Polarizing Intermediate Attachment (U-CPA)
# Always make sure to mount this unit in such a manner, that it is
parallel to the arm.
Intermediate Attachment for AN360P (U-OPA)
# The position of this intermediate tube can be adjusted later. At this
point, place the tube so the nameplate is toward the rear.
(Fig. 6)