TROUBLESHOOTING GUIDE

Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems
occur, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for
assistance.
Problem
Bulb does not light.
Bulb is blown immediately.
Field of view is dark.
Brightness cannot be varied.
Field of view is obscured or not
evenly illuminated.
Lamp
Cause
LED
Power is not ON.
Common
Power cord is unplugged.
Common
Connectors are connected im-
properly.
Halogen
Bulb is not attached.
Halogen
Bulb is burned out.
Halogen
The TL4 is not ON.
Halogen
The bulb is not a designated
one.
LED
LIM setting is inappropriate.
Common
Aperture and field iris diaphragms
are not opened wide enough.
Common
Light path selector knob is set to
the
position.
Common
Condenser is not properly cen-
tered.
Common
Condenser height is too low.
Common
ND filter is engaged.
LED
LIM is not ON.
Halogen
LED brightness adjustment knob
is rotated.
Common
Condenser is not properly
centered.
Common
Condenser height is too low.
Common
Light path selector knob is set in
an intermediate position.
Common
Objective is not correctly enga-
ged.
Common
Condenser top lens is not swit-
ched properly.
Common
An objective that falls outside
of the condenser's illumination
range is used.
Common
Condenser is not properly
centered.
Common
Field iris diaphragm is stopped
down too far.
Halogen
Bulb is not mounted correctly.
Remedy
Set the main switch to " I " (ON).
Plug power cord into the power
outlet.
Connect connectors securely.
Attach a halogen bulb.
Replace bulb.
Set the main switch of the TL4
to " I " (ON).
Replace bulb with a designated
one.
Set LIM again.
Adjust them to proper sizes.
Set the knob to the
position.
Center the condenser.
Adjust correctly.
Disengage the ND filters.
Switch the LIM function OFF.
Use the brightness adjustment
knob of the TL4.
Center the condenser.
Adjust correctly.
Set to the desired click position.
Make sure that the revolving
nosepiece clicks properly into
place.
Swing out the top lens when
using a 4X or higher-power
objective.
Use the objective with the specified
magnification.
Center the condenser.
Open the field iris diaphragm to
an optimum size.
Push the pins of halogen bulb all
the way until the stop position.