Olympus BX51M Manuel d'instructions - Page 13
Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus BX51M. Olympus BX51M 36 pages. System metallurgical microscope
@
Fig. 10
Field iris diaphragm image
Field of view of eyepiece
@
Fig. 11
70%
30%
Aperture iris image
2
Centering the Field Iris Diaphragm (FS)
1. Slide the mirror selector lever @ to "BF".
²
2. Engage the 10X objective by rotating the revolving nosepiece, place the
specimen on the stage and adjust approximate focusing.
3. Pull out the FS knob ² on the reflected light illuminator to reduce the
aperture iris diaphragm a little.
4. Rotate the two FS centering screws ³ using the Allen screwdriver to
³
adjust so that the field iris image becomes concentric with the field of
view.
5. While pushing in the FS knob ², open the field iris diaphragm until the
field iris image inscribes the field of view. If the image is found to be
eccentric, adjust the centering again.
6. Open the field iris so that its image is almost the same size as (i.e. sub-
scribes) the field of view.
Using the Field Iris Diaphragm
· In reflected light brightfield observation
The field iris diaphragm adjusts the illuminated area to obtain an image
with high contrast.
According to the objective in use, adjust the FS knob ² of the reflected
light illuminator until the iris image circumscribes the field of view to
block unnecessary light.
· In reflected light darkfield observation
The field iris diaphragm must be opened by pushing in the FS knob ².
3
Centering the Aperture Iris Diaphragm (AS)
1. Slide the mirror selector lever @ to "BF".
²
2. Engage the 10X objective by rotating the revolving nosepiece, place the
specimen on the stage and adjust approximate focusing.
3. Remove the eyepiece, look into the eyepiece sleeve and pull the AS
knob ² so that the aperture is about 70%.
³
4. If the center of the iris diaphragm is deviated, center it by rotating the two
AS centering screws ³ using the Allen screwdriver.
|
Using the Aperture Iris Diaphragm
· In reflected light brightfield observation, optimum observation is gen-
erally possible by setting the aperture to between 70% and 80% of the
aperture number of the objective.
· In reflected light darkfield observation, the aperture must be fully opened
by pushing in the AS knob ².
}With some specimens, an image with high contrast and little flare may
sometimes be obtained when the aperture is slightly closed. It is there-
fore recommended to also try a slightly closed aperture.
BX51M
(Fig. 10)
(Fig. 11)
10