Olympus BX53 Manuel d'instructions - Page 29

Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus BX53. Olympus BX53 48 pages. Polarizing microscope
Également pour Olympus BX53 : Manuel de l'utilisateur (9 pages), Manuel d'instructions (36 pages)

Olympus BX53 Manuel d'instructions
2
1
Fig. 42

5-5 Condenser

For how to adjust the centering of the condenser, see page 12.
1

Compatibility of Objectives and Condensers

Objective
Magnification
magnification
1.25X
2X
4X
Applicable
10-60X
100X
* When using the U-SC3 swing-out condenser together with a 1.25X to 4X objective, fully open the condenser aperture
iris diaphragm and use the field iris diaphragm in the base as aperture diaphragm. With the 1.25X to 2X objectives, the
surroundings of the field of view may become dark.
} To obtain better illumination in camera recording using a 1.25X to 4X objective, use of the U-ULC-2 is recommended.
**When the U-LC is used with an objective with an NA of 1.0 or more, the NA of the lighting becomes insufficient.
} When the U-LC is used, it is required to attach the auxiliary lens (page 35).
6

Using Eyepieces Incorporating a Micrometer

} When the eyepieces in use incorporate a micrometer, the accuracy of
the left-right focusing adjustment (diopter adjustment) can be improved
further.
1. Looking into the right eyepiece with your right eye, turn the top of the
eyepiece @ so that the micrometer in the field of view looks sharpest
(Fig. 42).
2. Looking into the right eyepiece, rotate the coarse and fine adjustment kn-
obs to bring the specimen and micrometer (cross lines, etc.) into focus.
3. Looking into the left eyepiece with your left eye, rotate the diopter adjust-
ment ring 2 to bring the specimen into focus.
Low
Abbe
U-AC2
U-LC
Applicable
(FN 22)
(FN 22)**
Applicable
(FN 26.5)
Condenser
Achromat/aplanat
Swing-out
U-AAC
U-SC3
Applicable
(FN 22)
Applicable (FN 26.5)
by moving top lens
out of the light path*
(FN 26.5)
Applicable
Engage top lens in
(FN 26.5)
light path (FN 26.5)
BX53
Ultralow
magnification
U-ULC-2
Applicable
(FN 26.5)
25