Olympus BX53 Manuel d'instructions - Page 17
Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus BX53. Olympus BX53 36 pages. Polarizing microscope
Également pour Olympus BX53 : Manuel d'instructions (48 pages), Manuel de l'utilisateur (9 pages)
Fig. 15
Fig. 16
4
Stage Height Adjustment
} By lowering the position of the substage, the microscope will
accommodate specimens with a maximum height of 35 mm. This is
useful when observing metallurgical specimens and other thick objects.
1. Lower the stage to its lower limit, then remove the stage from the
microscope. (See page 4.)
2. Using the Allen screwdriver, loosen the substage bracket clamping screw
@ and remove the substage assembly. (Fig. 15)
3. Turn the coarse adjustment knob and raise the focusing block ³ to
where the stopper screw ² in the arm becomes visible. (Fig. 16)
4. Using the Allen screwdriver, loosen and remove the upper stopper screw
².
5. Reattach substage assembly and stage.
} Store the remove stopper screw ² in a safe place so that it will not
be lost, if needed again.
BX53/BX53M-P
(Figs. 15 & 16)
14