Olympus CHS-F Manuel de réparation - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel de réparation pour {nom_de_la_catégorie} Olympus CHS-F. Olympus CHS-F 42 pages. Biological mlcroscope

Olympus CHS-F Manuel de réparation
3-7. Remove
back cover
(A8466300)
from
the
Arm (C2-AS).
3-7-l.
Pull out 6 plastic rivets (A8435500).
*Rivets
are set as indicated
in the following
sketch and will be easily released by pull-
ing on their heads.
3-8.
3-9.
Remove the rack for coarse focus adjustment
(A8469300)
by removing
2 screw (A83
x
IOSA) from the back of the arm.
Remove 4 screws (A83 x 10SA) from under-
neath the condenser
carrier (A8468800)
and
release the guide (A8468600)
and the block
(A8009700).
-