To reduce the flare, set a lower exposure using the exposure adjustment function or setting the exposure manually.
(Details of STM6 combination)
Edge observation using the MM6-OB3X/5X/10X objective in combination with the MM6C-VL/MM6-ETR
8. When the SZX16 is combined, the peripheral brightness may become insufficient with a zoom ratio of no more than 10 x.
9. The image of dark specimens under reflected light or in the darkfield (specimens that need exposure of 1/2 sec. or more
at ISO 100 equivalent) cannot be recorded. When recording the image of the particularly dark specimen, the lack of pixel
caused by the cosmic ray effect or the constant bright point may be noticeable, but they are not failures.
10. Specimens with a distribution that is not suitable for center averaged metering require spot metering or exposure adjust-
ment.
11. When electronic zoom is used for magnification display during focusing, the image may become noticeably coarse with
certain samples.
12. The live image frame rate of 15 fps is possible when the exposure time is no more than 1/15 sec.
13. The traceability of auto white balance control deteriorates when the specimen contains little white area.
14. We recommend you to use the same live resolution as the image quality mode during recording. If the live resolution is
different from the image quality mode, the measurement result may be shifted.
15. If the microscope light source is set too bright, color unevenness may occur in photographed images. Should this be the
case, adjust the brightness to an appropriate level (for example, by lowering the lamp voltage or by inserting a neutral
filter).
16. When combining to CX or CKX, set the light volume of the microscope illumination closer to the maximum level in order
to gain the best color reproduction.
17. To prevent destruction of the recorded image (data), never perform the following action during recording of a still image or
movie;
· turning power OFF;
· disconnecting the AC adapter;
· disconnecting the camera cable;