Olympus EHT-431 Manuel d'instruction - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Olympus EHT-431. Olympus EHT-431 24 pages.

2 )
Light Source
The light source consists of lamp house
CD
and lamp socket
® .
The lamp socket is provided with two coaxial lamp centering knobs
@ _
The Jamp socket can be moved back and forth along the optical
axis to eliminate uneven illumination in the field of view.
a.
Electric Connection
1) Insert the jacks of the lamp cord into the low voltage outlet at
the back of microscope base.
2) Insert the plugs of the line cord into the tine cord socket and the
AC
po......er
outlet respectively.
b.
Adjustment of Line Voltage
1) Turn the voltage adjustment knob clockwise to position ON.
2) If the bulb is dimmly lit. the line voltage is proper, and you have
only to manipulate the voltage adjustment knob in order to obtain
optimum light intensity. with no need to further proceed to the
following procedure 3).
3) Even after the voltage adjustment knob is turned on, if the bulb
does not light or lights up bright immediately after switching on,
rotate gradually the line voltage adjustment screw (at the back of
the base) with a coin, until the lamp dims.
o
According to the fluctuations of line voltage and frequency
(SO/60Hz), minimum voltage required for lighting the bulb
varies; if the bulb does not light at all or lights up immediately
after switching on, re-adjustment of the line voltage screw is
necessary to dim the bulb.
Note: For light intensity adjustment after dimming the bulb,
use the voltage adjustment knob on the side of the base.
c.
Attach the tamp socket.
1) Loosen the clamping screw
<!) .
2) Insert the lamp socket
@
into the lamp house
CD .
3) Tighten the clamping screw to secure the socket.
9