Operation
4
Caution
# When transporting the microscope, do not hold it by the revolving nosepiece.
# Be careful not to pinch your finger between the hand guard and the objective when rotating it.
# The parfocal property when the objective is replaced is not completely guaranteed.
23
...
Š
Fig. 31
‰
Fig. 32
Fig. 33
5. Place the nosepiece mount ... (with its objective mounts † facing
upward) where the objective has been by aligning the screw holes.
Using the Allen wrench (for M3 screws), clamp the revolver mount with
the provided 3 clamping screws (M3 –– 6 mm long). (Fig. 31)
# As the screw holes may be hidden behind the objective mounts †,
clamp the nosepiece mount while rotating it. (Fig. 31)
# Before installing the focusing assembly on the pillar, adjust the coarse
and fine focus adjustment knobs so that the hand guard Š does not
protrude from the lower end of the focusing assembly ‰. (Fig. 32)
6. Install the focusing assembly on the pillar, and install the observation
tube and intermediate attachment which has been removed above in
the original position.
7. Attach two objectives onto their respective objective mounts † by screw-
ing. (Fig. 31)
8. Attach the magnification indicator ring provided with the objectives on
the zooming knob (see page 10).
Hold the objective and gently rotate it until a click position where the
objective to be used is engaged in the light path.