Olympus SZX10 Manuel d'instructions - Page 14

Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus SZX10. Olympus SZX10 36 pages. Research high-class stereo microscope
Également pour Olympus SZX10 : Vue d'ensemble (24 pages), Manuel d'instructions (40 pages)

Olympus SZX10 Manuel d'instructions
|
WHSZ10X-H
10
Fig. 10
@
²
³
Fig. 11
When Using the Eyepiece Micrometer Disk
1. Look into the right eyepiece that includes the eyepiece micrometer disk,
and focus the micrometer disk by rotating the diopter adjustment ring ².
2. Place an easy-to-observe specimen on the stage plate.
3. Rotate the zooming knob ³ to the highest magnification and, looking
only into the right eyepiece, focus the specimen using the coarse and
fine focus adjustment knobs.
Ensure that both the eyepiece micrometer disk and specimen are fo-
cused accurately.
4. Rotate the zooming knob ³ to the lowest magnification, look only into
the left eyepiece, and focus the specimen by rotating the left diopter
adjustment ring | instead of the coarse and fine focus adjustment knobs.
}Note (or memorize) the diopter readings of the left and right eyepiece
scales so that they can be duplicated quickly in the next observation.
3

Using the Eye Shades

When Wearing Eyeglasses
Use with eye shades in their normal folded-down position. (This will
prevent eyeglasses from being scratched by the eyepiece.)
When Not Wearing Eyeglasses
Extend the folded eyeshades in the direction of the arrow. This makes
observation easier by preventing the inverse incidence of light from be-
tween the eyepiece and your eyes.
4

Using the Eyepiece Micrometer Disk

}A variety of eyepiece micrometer disks @ can be inserted into the
WHSZ10X-H, WHSZ15X-H and WHSZ20X-H eyepieces. Please pur-
chase an eyepiece micrometer disk with a diameter of 24 mm and
thickness of 1.5 mm.
1. Unscrew the micrometer disk frame ² by turning it counterclock-
wise to remove it from the bottom of the eyepiece.
2. Clean the eyepiece micrometer disk @ to remove dust and dirt, then
place the disk into the micrometer disk frame ² so that the side with
reticule ³ faces downward.
3. Attach the micrometer disk frame ² containing the eyepiece mi-
crometer disk @ by gently screwing it onto the eyepiece |.
# The WHSZ20X-H eyepiece is designed to apply a magnification to
the focusing plane of the micrometer disk.
As the magnification coefficient is 1.35, be sure to perform magnifi-
cation compensation when using this eyepiece in measurement.
When the micrometer disk is engaged in their light path, the light
path length will extend and the diopter scale may deviate toward the
+ direction from the normal indication. However, this does not pose
any problem in actual observation.
}When the eyepiece micrometer disk @ is not used, wrap it in a clean
sheet of soft paper before storage.
(Fig. 10)
(Fig. 11)