4

Zoom Magnification Indication

The body zoom magnification a is indicated on the zooming
knob on the right side.
The total magnification of observation can be calculated with the
following formula:
Objective
x
magnification
5
2X Objective Collection Collar
· The SDFPLAPO2XPFC objective is provided with a correction
TIP
collar a for use in compensating for the aberration caused due
to the observation medium such as water and plastic container.
When observing a specimen in a liquid or through a petri dish
cover, turn the collar to obtain the optimum contrast.
· Correction is possible for aberration corresponding to
water depth of about 5 mm.
· The correction effect may be less tangible when the
zoom magnification is low or the aperture iris diaphragm
is stopped down.
· When using the SZX2-2RE16 revolving nosepiece, hold
the objective during rotation of the ring so that the revolving
nosepiece is not deviated the click position.
Body zoom
Eyepiece
x
magnification
magnification
SZX16
9