Mounting the Microscope Body
1
Remove the cap a on the focusing assembly by inserting a thin
object into the notch.
2
Using the provided Allen wrench, loosen the dovetail mount clamping
screw inside the cap on the focusing assembly by rotating it by 2 or
3 turns (counterclockwise).
3
Gently insert the dovetail mount c on the rear of the microscope
body into the dovetail mounting port b on the focusing assembly.
Do not insert the mount at an angle or with excessive
NOTE
force, for this may cause malfunctions.
4
When the microscope body has been inserted until it stops, tighten
the clamping screw using the Allen wrench.
5
Place the cap a in the original position.
4
Mounting the Objective
Since the objective is heavy, take the following measures
NOTE
to prepare for an accident in case it is dropped.
· Attach the cap to the objective tip.
· Place the cap of the objective mount thread or a notebook,
etc. on the stand to absorb the shock even when the
objective drops.
While holding the tip of the objective b , mount it on the objective
mount thread a by rotating the objective in the direction of the arrow.
5
Mounting the Observation Tube
1
Using the Allen screwdriver, loosen the observation tube clamping
screw a completely.
2
Aligning the positioning pin c of the microscope body to the
positioning groove b on the observation tube, insert the dovetail
mount d at the bottom of the observation tube into the mounting
port e on the microscope body.
3
Using the Allen screwdriver, tighten the observation tube clamping
screw a .
The observation tube can be mounted 180° from the
NOTE
above position, but this positioning makes observation
difficult and is not recommended.
When the auxiliary pillar is used, this positioning is