A.
SEQUENTIAL PROCEDURE OF DISASSEMBLY
(See Fig. 1)
1.
Disassembly of Focusing Mechanism
1-1)
Rotate focusing knobs until main body goes up as well as possible.
1-2)
Unscrew
~038
that can be seen in dovetail block AA4869.
1-3)
Disengage metal block AA0738 from dovetail AA4869.
Rotate focusing knobs, and lower main body so that rack and pinion are
disengaged, and pull out
metal block.
1-4)
Remove focusing knobs AA0005.
Holding both of focusing knobs in both hands, rotate them in the oppo-
site directions simultaneously so that they are loosened.
When they are
completely loosened, focusing knob AA0005 and nut AA0008 are disengaged.
1-5)
Pull out pinion from bearing.
As you pull out focusing knob AA0006 in the axial direction, pinion can
be pulled out.
1-6)
Unscrew two CMK4x20SA and pull out block AA4869 from main body.
2.
Disassembly. of Main Body and Lens Drum
2-1)
Unscrew four CMK3xl0SA and pull out fixing guide AA4947.
2-2)
Unscrew two SK1.4-21, and remove cover plate AA4948.
2-3)
Remove three glass discs LP0387.
Prior to removal, immerse three glass discs with a little amount of
alcohol for a few minutes, because these discs were shellaced, then
stick them to Scotch tape to remove with ease.
2-4)
Unscrew two HK2x4SA, and remove click-spring AA4858.
2-5)
Unscrew left knob AA4861.
After removal of screw AA4866, insert a small screwdriver through empty
hole to loosen setscrew AA6207.
Knob can be removed.