8. Inspection label
No.
430
Cover or exchange an inspection label (as shown below) at the enclosure back of the unit and mark the
due date of the next periodic safety check (month/year). The unit must undergo a safety check at annual
intervals.
Documentation
Document the test results in "SCR.991007".
440
© Celon AG
,
Issue date: 01.09.2008
SCI.991007-1.4
Safety Check
Instruction.doc
Safety Check Instruction
Procedure
Page 9 / 9