Chapter 4 Operation
4.2
Remove water from the Air/Water Channel (For
endoscope with Air/Water channel only)
When Irrigating With 70% Ethyl or Isopropyl Alcohol
18
Do not remove the AW Channel Cleaning Adapter and the
Alcohol Container's nozzle until the Power Switch is OFF
"
Container's nozzle are removed from the endoscope while
the Power Switch is ON "
alcohol will splatter.
Flushing the channels with 70% ethyl or isopropyl alcohol
facilitates drying inside the Air/Water channel.
1.
Turn the Power Switch ON "
2.
Depress the button on the AW Channel Cleaning Adapter for approximately
10 seconds to irrigate the channel with 70% ethyl or isopropyl alcohol (see
Figure 4.1).
3.
Release the button and perform insufflation for approximately 10 seconds
until no more 70% ethyl or isopropyl alcohol exits the endoscope.
4.
Turn the Power Switch OFF "
5.
Remove the Alcohol Container's nozzle from the Air/Water Port on the Light
Guide Connector.
6.
Perform drying the Air/Water channel according to the next page.
Figure 4.1
MAINTENANCE UNIT MU-1
". If the AW Channel Cleaning Adapter or Alcohol
".
".
", water or 70% ethyl or isopropyl