3 Scope of supply

2.8

Conditions for placing, operating and storing

• The maximum permissible range for the ambient temperature during
• The maximum relative humidity: 80% at temperatures up to 31 °C, linearly
• Only use the system indoors and in a laboratory or laboratory-like
• Only use the system up to a maximum altitude of 2000 m.
• The electricity supply should not fluctuate by more than 10% of the
• Pollution degree: 2 (IEC60664)
• Installation category (overvoltage): II (IEC60664)
3
Scope of supply
System Setup Guide
operation of the CIX100 system is 5 °C to 35 °C.
decreasing to 70% at 34 °C, 60% at 37 °C, and 50% relative humidity at
40 °C.
environment.
nominal voltage.
1
CIX100 Microscope system
1
System Setup Guide
1
Joystick
1
Box with instructions for the individual components
1
Box of cables
1
Particle standard device with holder
1
Filter holder
1
Multi-sample holder
1
DVD with the CIX ASW software
1
Microsoft Office 2016 (installed and activated)
1
Computer
1
Touchscreen monitor
1
Keyboard and mouse
1
Calibration Protocol
10