Olympus MX61A Manuel - Page 24

Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} Olympus MX61A. Olympus MX61A 48 pages. Motorized 300 mm compatible semiconductor/fpd inspection microscope

12
Connecting the Module Connection Cables
Since this system includes motorized modules, do not connect the power cord until all other connections have
been completed.
Cables and cords are vulnerable to bend or twist. Do not apply excessive force to them.
Always connect the module designated by Olympus to each connector. Insert each connector in the correct
orientation and tighten the clamping screws if these are provided.
Use a PC that complies with the IEC60950-1 requirements. Otherwise, we cannot guarantee the performance of
the entire system.
# Always use a commercially available straight cable as the RS232C cable (use of an inappropriate cable may
result in failure). The cable should be equipped with D-sub 9-pin (female) connectors on both ends.
When the MX-OPU61A is used, the RS232C cable is provided.
Motorized
DIC Slider
U-DICRM
Computer
(running MX2-BSW software)
21
Motorized
Analyzer
MX-ANM
Control Box
BX-UCB
RS232C cable
} The MX-OPU61A
is not required if
a computer is
used.
Microscope
MX61A
RS232C cable
RS232C
RS232C cable
cable
Operation Unit
MX-OPU61A
Power Cord
Active AF Unit
U-AFA2M
Remote Cable
Rear view of the
U-RMT
base of MX61A
Unused
connector
Control
Board
U-ZPCB or
U-AFA2M-CB
To
U-UVF248LBM
Rear panel
of BX-UCB
Option Board
U-DICPCB
Power Cord
Hand Switch
MX-HS61A