Olympus U-AFA2M-DUV Manuel d'instructions - Page 10
Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus U-AFA2M-DUV. Olympus U-AFA2M-DUV 24 pages. Active autofocusing unit
NOMENCLATURE
Active AF Unit U-AFA2M-VIS/U-AFA2M-DUV
Be sure to use the connection cable designated by Olympus.
The computer in use should meet the IEC60950 requirements.
If any item other than specified is used, the entire performance of the system cannot be guaranteed.
Laser beam path A
Laser beam output
directions
Laser beam path B
AF Cable U-AFA2M-CBL2M (2 m)/U-AFA2M-CBL3M (3 m)
AF Control Board U-AFA2M-CB
}To be installed in the BX-UCB Control Box.
Termination connector
#If disconnected, AF operation
will be unavailable.
CAUTION
If another Z-focusing control board (U-
ZPCB, U-AFP1-CB, etc.) is already installed
in the BX-UCB, be sure to remove it before
installing this board.
Otherwise, malfunction will result.
6
Observation tube mount
(Centering adjustment type)
Warning label (P. 2)
AF board connector
Class 1 IEC60825-1:2007 / IEC60825-1:2014
Laser beam pathA:Below 1μW
Laser beam pathB:200μW Maximum
Warning label (P. 2)
Connection cable
U-AFA2M
BX-UCB (AF Control Board)
AF control board connector
The three other connectors
are not used.