Olympus U-RFLT50 Manuel d'instructions - Page 10

Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour {nom_de_la_catégorie} Olympus U-RFLT50. Olympus U-RFLT50 16 pages. Power supply unit
Également pour Olympus U-RFLT50 : Manuel d'instructions (36 pages)

Olympus U-RFLT50 Manuel d'instructions
3
3

PREPARATION FOR OBSERVATION

3-1 Turning ON the Mercury Burner

Fig. 9

3-2 Centering the Mercury Burner

Center the mercury burner as described in the respective instruction manuals provided with the CX-RFL, CH-RFL, etc.
reflected light fluorescence attachments.
5
1
Turning ON the Mercury Burner
# Confirm that the power supply voltage setting, frequency setting, and
burner type setting are correct. Improper settings may reduce the
burner life time considerably.
Press the power supply unit's main switch to " I " (ON). Between 5 to 10
minutes are required for the arc to stabilize after the burner is ignited.
# Some mercury burners may not ignite the first time the power is turned
1
ON. If the burner does not ignite, the power supply unit's starter safety
device will activate and should be reset. Press the main switch to
"
" (OFF) once. Then insert the tip of a mechanical pencil, etc. into
the starter reset switch hole @ on the right hand side of the power
supply unit and press the switch inside. Then press the main switch
to " I " (ON) again.
# To avoid shortening the life of the burner, do not turn the burner off
within 15 minutes of ignition. Use the shutter function to block light
if required.
# After turning the burner off, it cannot be re-ignited before the mercury
vapor cools and condenses to a liquid. Wait for about 10 minutes
before restarting the burner.
# The power supply unit has a limited operation life. Avoid burner
ignition switch operation when no burner is installed and do not
repeatedly turn on/off the burner with short intervals.
If the lamp housing is accidentally opened while the burner is
ignited, the safety interlock will activate and switch off the power
automatically. In this case press the main switch to "
wait at least 10 minutes before turning the burner ON again. Always
leave to cool before opening the lamp housing.
Make sure that the burner life time hour counter is properly reset to
"0.0" when a new burner is installed. In order not to impair the safety
of the equipment, replace the burner when the counter shows "100.0".
First press the main switch to "
10 minutes until the lamp housing has cooled down. The mercury
burner contains high-pressured gas and the burner may explode if
used beyond the specified life time.
" (OFF) and
" (OFF) and then wait for at least