Olympus UHI-4 Manuel de référence rapide
Parcourez en ligne ou téléchargez le pdf Manuel de référence rapide pour {nom_de_la_catégorie} Olympus UHI-4. Olympus UHI-4 4 pages. Insufflator
Disclaimer: Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each
item of equipment which should be thoroughly read and understood prior to handling, use and storage of the equipment.
Insufflator Controls
Power Switch
Set Pressure
Smoke Evacuation
CO
Connector
2
Connector
Issue 1
05/02/2021
UHI-4 Insufflator Quick Reference Guide
Actual Pressure
Volume Indicator
Flow Rate
Smoke Evacuation
Start/Stop
Speed
Cavity Modes
There are two cavity modes Normal and Small which regulate the pressure and flow rates.
Cavity mode
NORMAL
SMALL
Press to toggle between normal and small
Note:
The cavity mode cannot be changed once insufflation has started, if required stop
insufflation and then change mode.
Flow Rates
The are three flow settings:
Cavity mode
NORMAL
SMALL
Press to toggle between flow rates
Warning:
when using a Veress needle always use the LOW flow mode as insufflation in other
modes is dangerous because the cavity pressure may exceed the set pressure.
Setting pressure range
Note
To increase the pressure from 20
3 - 25 mmHg
to 21 mmHg, keep pressing "+"
(0.4 - 3.3 kPa)
for at least three seconds
3 - 15 mmHg
(0.4 - 2.0 kPa)
Setting pressure range
Note
HIGH
20 - 45
MED
1.1 - 1.9, 2 - 5
LOW
0.1 - 1.0
HIGH
6 - 10
MED
1.1 - 1.9, 2 - 5
LOW
0.1 - 1.0
KMF-08929
NCDR 02867