AccuLine 40-6680 Manuel d'entretien - Page 15
Parcourez en ligne ou téléchargez le pdf Manuel d'entretien pour {nom_de_la_catégorie} AccuLine 40-6680. AccuLine 40-6680 16 pages. 5 beam laser pointer
Également pour AccuLine 40-6680 : Manuel d'instruction (13 pages)
4.4. Front, Left, and Right Laser Calibration
3.
When the front, left and
right laser points are
beyond tolerance as
shown in the figure to
the right, you need to
adjust the screws
shown in the figure
below.
When the
instrument is in the
state7, you can
loosen screw 7 and
tighten screw 8.
When the
instrument is in the
state 8, you can
loosen screw 8 and
tighten 7.
When the
instrument is in the state 9, you can loosen screw 6 and tighten 5.
When the instrument is in the state10, you can loosen screw 5 and tighten 6.
4.
After the above adjusting is completed, double check accuracy of the entire instrument as described in 4.1. If there still is
error, repeat respective calibrations until the unit is qualified.
Prepared By: Tim Wojo
Laser
Circle with a diameter of
Reference point
State 7
Laser spot
Circle with a diameter of
Reference
State 9
Screw 6
Revision - Date: 0 – 20080430
40-6680 Service Manual
Circle with a diameter of
Reference point
Laser
State 8
Circle with a diameter of
Reference
State 10
Screw 7
Screw 5
Screw 8
Page 15 of 16
Laser spot