Arbutus Medical DCB127 Manuel d'utilisation - Page 4
Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Arbutus Medical DCB127. Arbutus Medical DCB127 18 pages.
1. WARNINGS
Solutions that contain or release chlorine (eg. chlorine bleach, Presept
by J&J) or hydrogen peroxide will damage the product. Refer to the
instructions in Section 7: Reprocessing.
Handle with care. Dropping the device on a hard surface can
permanently damage the sterile barrier. Follow guidelines for
preventing mechanical damage outlined in Section 5.1. Inspect device
after each use following procedure outlined in Section 5.3 and replace if
integrity has been compromised. Insert a strong, cool light inside of the
Drill Cover linen to inspect for damage.
Drilling of bone can produce heat at the cutting site and damage
surrounding bone. Control bone temperature. Drill only in short bursts
and use irrigation.
The device is not suitable for use in oxygen rich environments. Ensure
staff have been trained on surgical fire prevention associated with
electrical equipment, and have a plan of action in the case of fire.
For electrical safety, consult the enclosed Dewalt tool manual. This
device generates electromagnetic interference within approved
limits for medical devices. If interference were to occur in a particular
application, stop operation of the device. Reorient or relocate the
affected equipment. There is no significant risk of reciprocal interference
posed by the device during specific investigations or treatments.
Do not oil or lubricate unless with biocompatible lubricant.
For more information regarding these warnings, contact Arbutus Medical.
Arbutus Medical
4