4. STERILE LOADING PROCEDURE
The following step-by-step procedure describes the sterile loading process for the DrillCover
Hex System. Two team members are required for loading. Observe when a sterile or
nonsterile team member should perform each step.
Arbutus Medical recommends practicing the sterile loading procedure before clinical
use.
Complete the loading procedure before the sterile team member is contaminated by
biohazard during the procedure.
4.1. Obtain Sterile Equipment
Sterile Team Member
1. Collect sterile parts, including: DrillCover
linen, chuck adapter, and chuck key.
2. Check parts for cleanliness.
3. If soiled, send soiled components for
reprocessing and use a cleaned and
sterilized DrillCover.
DrillCover Hex System: Instructions For Use