Common Error messages
Message
no Enc
no Pos
b fail
f fail
P fail
LOCK
P LOCK
no oFF
MON flashing
no Co
uL, LL
Manual Part # 800-1403-001 Revision 06-01-2020
There is not an encoder connected to the readout, or the
connected encoder has a fault.
The readout has not been calibrated yet.
The battery in the readout needs to be replaced
immediately.
The readout cannot display fractional values larger than
399 63/64 inches.
The battery in the readout needs to be replaced
immediately. Check programming values after new battery
is installed.
The keypad is LOCKed, but an attempt was made to
change the calibration. Unlock keypad to fix, see page 6.
The keypad is LOCKed, but an attempt was made to
change the calibration. Change Pr3 to fix, see page 11.
Offsets are disabled, but an attempt was made to apply an
offset value.
Monitor mode is turned on, and the system is out of the
allowable position.
Non-linear compensation is enabled, but there is no look-
up data for the displayed measurement.
Limits mode is turned on, and a limit has been exceeded.