Creche Innovations MicroCool Manuel de l'utilisateur
Troubleshooting

Troubleshooting Guide

Symptoms
No Light Output
Poor (low) Light Output
Light head does not hold
position
Probable Cause
Not plugged in.
Switch not "on" (I or II).
LED failed.
Fuses failed:
Because of internal short.
Because of high mains
Voltage, or high voltage
spikes.
Because of single
Occurrence of lightening
strike.
Break in wiring.
Short or open circuit in
power supply.
Switch failure
Loose or corroded
connectors.
Faculty LED module.
Weak arm
Lever not tightened
enough
Worn or missing
components.
10
Remedy
Plug into a suitable
hospital-grade grounded
socket.
Turn switch on (I), left or
right.
Replace with new LED
module.
Replace with new fuse(s).
Locate short and fix it.
Put voltage regulator
on line.
Check condition of
transformer and wiring.
Locate and replace wires.
Determine failure mode
and replace transformer.
Replace switch.
Check all connections;
clean/replace as
necessary.
Replace power supply.
Replace LED module.
Replace arm
Tighten lever as required.
Replace as necessary.